Instructions for Use

Tray Liner: Best Practices for Use

PRIMED single use tray liners are intended to line the bottom of sterilization trays to absorb excessive moisture. They may also be used to cushion medical devices and protect sharp instruments. The product size is 20" x 25".

Reference

Product information

Indications for use

PRIMED tray liners are validated for use in pre-vacuum steam and EO sterilization systems:

  • Pre-vacuum steam at 270°F/132°C for 4 minutes.
  • 100% Ethylene oxide (EO) concentration of 725-735 mg/L at 131/55°C and 40-80% humidity for 60 minutes.

‍

Warnings

Not suitable for use in dry heat, hydrogen peroxide and gas plasma sterilization systems.

Disclaimer

The information supplied in this document is for guidance only and should not be construed as a warranty. All implied warranties are expressly disclaimed, including without limitation any warranty of merchantability of fitness for use. All users of the material are responsible for ensuring that it is suitable for their needs, environment and end use. All data is subject to change as PRIMED deems appropriate.